Environmental Health & SafetyPotable (Drinking) WaterSan Diego State University Appendix A: Methods for Total Chlorine Residual Measurement Methods for determining total chlorine include amperometric titration, colorimetric DPD, titrimetric DPD, and iodimetric titration. The most widely used method for field test siturations is colorimetric DPD. In this method, DPD (N,N-diethyl-p-phenylenediamine) is oxidized by chlorine causing a magenta (red) color. The intensity of the color is directly proportional to the chlorine concentration in the sample. Colorimetric DPD field test kits are available from several suppliers. The basic procedures for the DPD Colorimetric method include the following steps:
Each chlorine test kit is designed to measure total residual chlorine only within a certain range of concentration. The low range instruments typically measure total residual chlorine in concentrations no higher than 3.5 to 5 mg/l. Some manufacturers have produced high range test kits that are capable of measuring chlorine at the level of doses required for disinfection of water mains, e.g. 25 mg/l. However, a low range test kit can be used to measure a chlorine concentration higher than the kit’s range by dilution. This dilution method can be performed by applying the graduated cylinder method as described below: Graduated Cylinder Method
(Vol. of distilled water + chlorinated sample) = Dilution Factor Where 2 ml of sample is combined with 48 ml of distilled water in graduated cylinder, the dilution factor is 25 as determined below: (48 ml of distilled water + 2 ml of sample) = Dilution Factor of 25 For example, if it is determined that the diluted sample from the graduated cylinder has a chlorine residual of 1 mg/ml, the undiluted sample from the disinfected main would have a residual of 25 x 1 mg/l or 25 mg/l. If it is not possible to accurately determine the chlorine residual of the undiluted sample, it may be necessary to provide a different dilution to the sample. For example, if it is anticipated that the chlorine residual is around 100 mg/ml (slug method), a more appropriate dilution factor would be 50. This could be obtained by diluting 1 ml of sample with 49 ml of chlorine free water. |
|
This page last updated September 24, 2008 |
||